Skip to product information
1 of 1

NWSPEC

ICP10

ICP10

Regular price $0.00 USD
Regular price Sale price $0.00 USD
Save Liquid error (snippets/price line 136): divided by 0% Sold out

Description:

ICP10 maintains rapid Analysis high sensitivity, wide dynamic range and high stability,achieves smaller device size and higher integration. It also adds a new inflation device to the light chamber, making the power on/off process more convenient. The operation is simple and the analysis is precise, providing users with an efficient solution for sample elemental analysis.

Key Features:

Rapid Analysis :

The speed of element detection is 15 elements per minute.The analysis time for each sample is either 2 minutes or 6 minutes. For the full-spectrum direct reading type, one sample can generally be analyzed in 2 minutes.

High Sensitivity: 

Utilizing advanced optical detection system and efficient plasma light source design, it can detect element concentrations as low as ppb (one billionth) level, meeting the requirements for high-precision analysis.

Wide Dynamic Range: 

The equipment has a wide dynamic range, capable of simultaneously detecting high concentrations and trace elements, providing a one-stop analysis solution for complex samples.

Simultaneous Analysis of Multiple Elements:

The ICP spectrometer supports simultaneous detection of multiple elements, with fast analysis speed and significant improvement in experimental efficiency.

Automated Control: 

With an integrated intelligent operating system, it automatically adjusts flow, gas, and power, simplifying operation steps and reducing manual intervention.

Smaller Volume: The excellent integration design enables the ICP10 to further reduce its volume, providing convenience for your experimental environment.

Durable Design:

The latest addition of an air chamber inflation device ensures that the precise optical components are less prone to moisture, while using high corrosion-resistant materials to ensure stable performance of the instrument over long-term use, suitable for various complex chemical sample analysis.

Easy Maintenance:

The modular design facilitates equipment maintenance and component replacement, reducing downtime and improving laboratory efficiency.

 Applications

Environmental monitoring, Mineral analysis, Metal smelting, Food safety,

Meeting the needs of all industries.

 Specifications

Dimensions (w x d x h  in cm)    

150 x 50 x 107                 

Weight (kg)  

235

Power

Single phase, 4-5.5 kW, 28 A, 220/240 V, 50/60 Hz

Spectrometer

1-m Czerny-Turner monochromator, argon purged, thermostated

Grating

configured upon request I: 2400 lines/mm, ion-etched holographic master, 80 x 110 mm

Option II: 4320 lines/mm used in first order

Option III: 3600 lines/mm used in first order

Option IV: 2400 lines/mm used in first order

Optics

Wavelength range

I: 160 - 800 nm

II: 160 - 530 nm

III: 160 - 442.5 nm

IV:  160 - 800 nm

Wavelength drive

Wavelength indication error ≤ ±0.03 nm, reproducibility ≤ 0.003 nm

Resolution

7/9/10/16 pm for configuration I/II/III/IV

Detector

Photomultiplier Tube (PMT)

Detection mode

Single channel

Read out

Digital display

ICP generator

Vacuum tube RF generator, frequency stabilized; reflected power control

Frequency (MHz)

40.68

Power (Watt)

4000-5500

Load coil

water cooled

Torch

fully demountable

Argon flow  (L/min)  

11.5-12.5

Sample introduction

thermoregulated sample compartment;  radial plasma orientation; Quick-release torch clamp

Operating system and software

Compatible with Windows 7, Windows 10, and Windows 11

Computer

Intel(R) Celeron(R) G6900, 8 GB DDR4, 4 MB Intel® Smart Cache, 3.40 GHz, Intel® UHD Graphics 710 keyboard, mouse, laser printer

Short term precision (RSD for 10 replicates)

 ≤ 1.5%

Long term precision (RSD for 4 hours)

 ≤ 2.0%

Sensitivity

 ppb level

Detection limits (mg/L):

Zn 213.856 nm ≤ 0.003

 

Mn 257.610 nm ≤ 0.002

 

Ba 455.403 nm ≤ 0.001

 

Ni 231.604 nm ≤ 0.01

 

Cr 267.716 nm ≤ 0.007

 

Cu 324.754 nm ≤ 0.007

Minimum spectral bandwidth Mn 257.610 nm

Better than national Class A standard Mn 257.610 nm spectral bandwidth (FWHM): ≤ 0.015 nm

Accessories

Quartz Torch for Oil Analysis, Oil-Cooled Spray Chamber, Quartz Nebulizer for Oil Analysis

 

 We can offer a full range of product training and method development services.

View full details